DC-STIMULATOR PLUS for tDCS and tACS

更新:2018/6/13 1:22:59      点击:
  • 品牌:   
  • 型号:   
  • 在线订购
产品介绍

The DC-STIMULATOR PLUS is a CE-certifiedmedical device for conducting non-invasive transcranial direct-currentstimulation (tDCS) on people. DC stimulation is used to treat strokes, epilepsy,migraine, tinnitus, depression, multiple sclerosis, dementia and chronicheadache.

The DC-STIMULATOR PLUS is a micro-processor-controlled constant currentsource. It meets the highest safety standards thanks to (hardware- andsoftware-based) multistage monitoring of the current path. By continuouslymonitoring electrode impedance it can detect insufficient contact with the skinand automatically terminate stimulation. This is a reliable method of excludingany injury to the patient.

Besides transcranial direct current stimulation (tDCS), the DC-STIMULATORPLUS also makes it possible to conduct bi-polar stimulation (tACS).

DC-STIMULATOR PLUS是一个CE认证的医疗设备,对人进行非侵入性的经颅直电流刺激(TDCS)。

DC-STIMULATOR PLUS是一个微型处理器控制的恒定电流源。它符合最高的安全标准(硬件和软件)的多级监控的当前路径。通过连续监测电极的阻抗,它可以检测到与皮肤的接触不充分,和自动终止的刺激。排除对患者造成伤害的任何可能性。

除了经颅直流电刺激(TDCS),DC-STIMULATOR PLUS 还使得能够进行双极刺激(TACS)。

DC-STIMULATOR PLUS features:

·        1 channel, unipolar (DC) and bipolar (AC) stimulation

·        Adjustable range of current ±4,500 µ, maximum adjustable currentstrength ±3,000 µ (peak-to-peak)*

·        Adjustable application time up to 30 minutes *

·        Frequencies adjustable up to 250 Hz, phase freely adjustable

·        4 standard modes - single (continuous stimulation) - pulse (cyclicalstimulation activation/deactivation) - sinus (sinus wave) - noise (normallydistributed)

·        customer-specific programs possible (optional)

·        "Study mode" for blind processing of genuine and 'pseudo'stimulation (optional)

External trigger input and output (optional)



1.     DC-STIMULATOR PLUS特点


● 微处理器控制的恒流源


● 1通道,单极(DC)和双极(AC)刺激


● 高安全标准,电流通路多极监控


● 刺激模式:单刺激(连续刺激,可调节,淡入和淡出)


● 刺激模式:脉冲(循环开启/关闭刺激,可调脉冲宽度和间隔)


● 刺激模式:正弦(双极正弦波,偏移、频率、相位和振荡周期可调节)


● 刺激模式:噪声,噪声LF(低频),噪声HF(高频)(正态分布宽频带的、低高频噪声,偏移和持续时间可调)


● 用于真刺激和伪刺激“盲”操作的学习模式,通过200个代码的编码表进行编码,独立可调设置(可保存,以避免对学习参数的偶然修改)


● 患者模式:简易(用于患者在家时的安全和受控使用,无参数修改的可能性)*


● 外部触发输入*,触发输出*


2.     DC-STIMULATOR PLUS规格


● 可调电流(DC)达4,500μA** 增量25μA


● 内部12bit D/A转换


● 内部时间分辨率1 ms(采样率1,280 sps)


● 最大1% 相对直流容错


● “单”刺激模式:持续时间15-1,800s **,增量15 s,淡入/淡出持续时间1-120 s,增1 s


● “脉冲”刺激模式:完整脉冲周期持续时间/刺激间隔(ISI)300-2,000 ms,增量100 ms,脉冲宽度200-(ISI-100),增量100 s,脉冲周期数1-500


● “正弦”刺激模式:可调电流达3,000μA (p-p)**,25μA增量,偏移0-±1,000μA,增量10μA,频率达250 Hz,增量0.01 Hz,可调相位0-360°,以5°的幅度,可调应用时间达30 min*


● “噪声”、“噪声LF”、“噪声HF”刺激模式:可调电流达±1,500μA (p-p),偏移0-±1,000μA,增量50μA,持续时间0-1,800 s**,增量5 s,电流在0-120 s期间内可调为达到或离开振荡水平


● 最大电压限值±20 V


● 内置可充电电池供电


● 连续操作时间约为6 h @1mA


● 字母数字显示器,带背光灯,4个按键的薄膜式键盘


● 接触保护的电极接头,符合DIN42802-2 (Φ1.5 mm)


● 耗电量约为1.2 W(取决于显示器亮度和应用电流)


● 尺寸:13.5cm×22.5cm×5.5.cm(W×D×H),重量(包括电池):0.8kg


*可选


**超过2,000μA的电流及多于20 min的应用时间只作研究用途。


DC-STIMULATOR PLUS的fMRI附加组件


● 在fMRI中使用DC-STIMULATOR


● 在EPI序列期间不会干扰fMRI图像


DC-STIMULATOR PLUS可选件


● 触发模块,可与外部触发器安全连接


使用正弦刺激模式时的相位同步触发输出